MEDIWISH · Cleaning verification
Test-soil removal at the selected load position
MEDIWISH Load Check is a single-use indicator for trained personnel monitoring cleaning in instrument washer-disinfectors. It shows whether the applied test soil has been removed at the selected location within a load. Supply is 100 complete cards per box, REF 2345678; use the reusable holder 12-5815.
- REF
- 23456782345678
- Pack
- 100 cardsPer box
- Card size
- 100 × 50 mm0.30 mm thick
- Holder
- 12-5815Reusable, ordered separately
What the indicator assesses
The card carries a proprietary test soil containing protein, lipid and polysaccharide components. The visual result concerns removal of that soil from the card in its defined holder and position. It does not establish cleanliness of every instrument, thermal disinfection or sterility, and does not replace other required load checks.
Install and inspect the complete card
The Load Check IFU, MW-LCI-IFU-001 Rev. 02, describes four test areas: both ends on both sides of the card. Holder 12-5815 provides two front and two aligned rear grid windows. Seat the complete card fully and align every test area with its window. Keep all four windows clear and attach the holder securely to a suitable basket rim at the position specified by the facility’s established procedure.
Use the cycle, chemistry, load, positions and frequency established for the actual process. After the cycle, follow precautions for hot loads and withdraw the card without washing, wiping or scraping it. Examine the actual surfaces on both sides against a clean white background under suitable lighting; the semi-translucent PP can show a mark from the opposite side.
Read and record the result
Under the Load Check IFU, PASS requires no visible test-soil residue in any of the four areas. Any visible residue, including a faint mark or small spot, is a failed cleaning-monitoring result. A lost, damaged, incorrectly positioned or manually wiped card, an interrupted cycle or another condition preventing reliable reading is invalid; do not record a pass. The PP may remain hazy or semi-matte: optical transparency is not the endpoint.
Follow the facility’s deviation procedure for failed or invalid results. Record REF, lot, holder, position, cycle, all four areas and corrective actions. Keep the affected load from further release pending assessment under that procedure.
Product specifications
| Characteristic | Specification |
|---|---|
| REF | 2345678 |
| Format | Single-use test-soil card |
| Carrier | Polypropylene (PP), semi-translucent, semi-matte, textured |
| Nominal dimensions | 100 × 50 × 0.30 mm |
| Pack | 100 complete cards per box |
| Compatible holder | 12-5815; ordered separately |
| Test configuration | Four inspection areas: both ends on both sides, per IFU Rev. 02 |
| Storage | 0–30 °C; 30–60% RH; closed original packaging, protected from moisture and direct sunlight |
| Shelf life | 24 months from manufacture under the stated storage conditions; observe labelled expiry |
Product · Samples · OEM enquiries
Ordering and documents
For a quotation or samples, specify REF 2345678, 100 cards per box and whether holder 12-5815 is required. The indicator is single-use; retain the reusable holder. Request the current controlled IFU, result guide and technical data sheet for your order. Reference instructions: MW-LCI-IFU-001 and MW-LCI-HIFU-001, Rev. 02, 8 September 2026.