The Mediwish® Three-Level Sequential Integrating Indicator is a chemical indicator for sterilization processes. It reacts within loads exposed to steam in a prevacuum sterilizer chamber, providing a reliable indication of time, temperature, and steam for safety. It has three levels of indication, with Level 2 confirming calibrated safety conditions and Level 3 indicating an overreach of integration conditions. It is a Class 5 chemical indicator with specific technical data and features such as traceability and error elimination. It is a cost-effective solution for sterilization cycles.
Description chemical indicators for sterilization
Ensure the utmost safety and efficiency in your sterilization processes with Mediwish® Three-Level Sequential Integrating Indicator. Our innovative autoclave chemical indicator is designed to react within loads exposed to steam prevacuum sterilizer chamber, providing a reliable indication time, temperature, steam of the process's safety.
At Level 1 - Process Exposure, the indicator turns black, signaling exposure to the sterilization process, though it doesn't assure the conditions for safety. Move to Level 2 - Correct Integration, where the indicator turns black, confirming that the calibrated conditions for safety have been met, ensuring a SAFE condition. Level 3 - Extended Integration indicates an overreach of integration conditions, signaling an EXTENDED condition.
Class 5 Chemical Indicator for sterilization specification:
Size strip: 20mmx140mm
Sterilization method: Steam
Initial indicatot color- blue, signal color dark brown
classification indicator: Type 5 (class 5) - class 5 integrating indicator (chemical integrators)
Ink: water based non toxic ink
Category: Internal chemical indicator, equal to biological monitoring
Class 5 integrator Technical data:
- PROCESS EXPOSURE STEAM CHECK: Indicates exposure to sterilization cycle monitoring process,
- Critical parameters: Exposure time: 7 min.; exposure temperature: 135C
- CORRECT INTEGRATION: Confirms calibrated safety conditions met (SAFE).
- EXTENDED INTEGRATION: Signals overreached integration conditions (EXTENDED).
- TRACEABILITY: Enables traceability of sterilization process information.
- ERROR ELIMINATION: Eliminates batch number recording errors.
- EASY INTERPRETATION: Type (class) 5 indicator ensures accurate color change for easy interpretation.
- COST-EFFECTIVE: Very low-cost solution for every sterilization cycles.