MEDIWISH · Cleaning monitoring guide
Read the result, then investigate
A failed cleaning-indicator result means the result did not meet the acceptance criterion for that indicator in the monitored configuration. It does not identify a single root cause. The investigation must consider the test assembly, the cleaning process and the way the result was read.
Separate visible residue from an invalid test
Read the complete working area using the model’s current instructions. MEDIWISH’s Rev05 specification distinguishes removal of the applied test soil from visible soil remaining. The Load Check result guide treats visible residue, including faint marks or small spots, as unacceptable. Do not invent a percentage-removal allowance or average a marked area with a clear area.
A lost, damaged, incorrectly positioned or manually wiped indicator may not provide a reliable reading. An interrupted cycle also needs to be handled as an invalid test under the applicable procedure. An invalid observation is not a pass. For Load Check, the matte or hazy appearance of the polypropylene carrier alone is not the test-soil endpoint.
Preserve the evidence before investigating
Do not rinse, wipe or scrape the indicator to improve its appearance. Record the REF, lot, cycle, holder, position and observation. Photograph the result when required by the facility’s procedure and retain the cycle record. Manage the affected load under that procedure; the Load Check IFU calls for holding further release pending assessment after a failed or invalid result.
Review possible contributors without guessing the cause
| Area to review | Evidence to compare |
|---|---|
| Indicator condition | Correct REF, expiry, packaging, storage and condition before use. |
| Installation | Correct holder, secure retention, orientation and unblocked access compared with the approved placement record. |
| Programme and load | Actual cycle record, interruptions, alarms and load arrangement against the approved configuration. |
| Cleaning chemistry | Correct product and evidence of dosing against its instructions and the facility procedure. |
| Water and equipment | Applicable water-supply checks, spray access and equipment condition assessed by responsible staff. |
| Reading | Complete area inspected under suitable lighting, without confusing the carrier with residual soil. |
Do not infer the exact detergent concentration or a specific equipment fault from the appearance of one strip. Likewise, the temperature of a later thermal-disinfection stage does not describe the actual washing conditions.
Link corrective action to the repeat test
Escalate the observation, correct identified causes and repeat monitoring with a fresh indicator according to the facility’s procedure. Reprocess the load where required. Preserve the original failure or invalid result and link it to the corrective action and repeat; a later clear indicator does not erase the first observation. The final load decision must include every required check.
Confirm the model and reading method
For model-specific interpretation, review MEDIWISH Load Check REF 2345678. Use the cleaning-indicator catalogue to confirm that the indicator and holder match the intended process.