A suitable cleaning indicator is one that can be used and interpreted within your washer-disinfector process. Start with the intended process and the manufacturer’s instructions, then compare the test-soil carrier, holder and evidence for that configuration. Colour and product name alone do not establish how difficult a soil is to remove.
Define the question before selecting the test
A ready-to-use soil-removal indicator shows what happened to its applied test soil at the monitored location. A residual-protein test examines material recovered from a sampled instrument surface. These are different observations. A cleared indicator does not establish that every instrument is clean, that thermal disinfection was achieved or that a load is sterile.
Write down the washer model, programme, chemistry, load arrangement and purpose of monitoring. Include the instruments’ processing instructions and the facility’s approved procedure. This gives the technical reviewer a defined configuration to assess instead of a request for the “strongest” strip.
Compare the whole indicator system
| Selection point | What to verify |
|---|---|
| Documented process | The specific REF is intended for the process being monitored. |
| Carrier and test soil | Card, polymer strip, steel coupon or complete device; the reading method applies to that model. |
| Holder | Correct REF, retention, access to the test area and approved orientation. |
| Evidence | Results from relevant cleaning conditions, including the limits of the assessment. |
| Routine handling | Pack size, lot traceability, expiry, storage and disposal can be managed reliably. |
MEDIWISH’s Rev05 specification identifies Load Check REF 2345678 with holder 12-5815; CEI CVC-50 and Universal Strip 647913 with holder 12-5814; and X-Soil Card 647914 with holder 12-5814-1. CVWC-20 has an integral holder and the complete device is single-use. A stated pairing is not permission to substitute one holder or process for another.
Interpret “challenge level” carefully
Two indicators can respond differently because of their soils, carriers, holders and exposure. A red variant is not automatically more challenging than a blue one. Ask for comparative results under a defined programme before ranking models. A 2025 primary study found process-dependent responses among eleven commercial indicators. It supports evaluating selection experimentally; it did not test MEDIWISH products. See Zhou et al., PLOS ONE.
Make selection reproducible
Retain the selected REF, compatible holder, instructions revision, approved locations and interpretation method in the monitoring procedure. The responsible team should define frequency and the number of test assemblies for the actual configuration. Changing the holder, chemistry, load or programme can require reassessment; moving an indicator to make it clear defeats a consistent comparison.
Compare the documented options
Use the cleaning-indicator and holder catalogue to compare formats. For a documented washer-disinfector card system, review MEDIWISH Load Check REF 2345678 and request the current model-specific instructions with your quotation or sample request.