MEDIWISH · Cleaning monitoring guide
Keep the result and its context together
A useful cleaning-indicator record connects an observation to the actual product, cycle and test position. A line saying only “passed” cannot explain which load was monitored or support an investigation later. Build the record so another trained person can reconstruct what was used, where it was placed and what happened next.
Keep identification and observation together
| Record group | Fields to include |
|---|---|
| Context | Facility, department, date, time and operator. |
| Cycle and load | Washer ID, programme, cycle number, load or basket ID and cycle-record reference. |
| Indicator | Product name, REF, lot, expiry and applicable instructions revision. |
| Installation | Holder REF and ID where used, position, orientation and placement-procedure revision. |
| Conditions | Applicable chemistry and dosing information, together with the source record. |
| Observation | Condition of each required test area, overall result and photograph reference where required. |
| Follow-up | Deviation, escalation, corrective action, load disposition, repeat-test reference and responsible review. |
Use the reading terms and test-area layout from the applicable model instructions. A result that cannot be assessed reliably must remain distinguishable from an acceptable result. Keep the indicator observation separate from instrument inspection, residual testing, thermal-disinfection records and the final load-release decision.
Preserve the original result
Document the first observation before a repeat test. Do not overwrite a failure with the later result. Give the repeat its own identifier and link it to the original record and corrective action. If a photograph is used, identify the REF and observation clearly and retain the original file under the facility’s record procedure.
For product traceability, use the actual label and issued documents. Do not generate replacement lot numbers, dates or regulatory identifiers. The current confirmed MEDIWISH Load Check pack is 100 cards; historical documentation referring to 200 cards should remain identifiable as historical and be reconciled through document control.
Compare results that answer the same question
Separate records by washer, programme, indicator REF, holder configuration and approved test position before interpreting a trend. Flag changes in chemistry, load arrangement, instructions or test location. Keep failures and invalid observations visible as different categories.
If you calculate an unacceptable-result proportion, state both the numerator and denominator. Do not silently exclude invalid tests; report them separately with the total tests recorded. A period with one failure in a small number of tests is not directly comparable to the same count across a much larger workload. No universal alert threshold or review interval is established here.
Use the trend to direct review
A pattern concentrated at one position may justify reviewing installation or access; a pattern across programmes may justify a broader investigation. These are investigation leads, not proof of causation. Check the underlying records and have the responsible team decide the response under the approved procedure.
Keep product identification consistent
Obtain the applicable record documents with MEDIWISH Load Check REF 2345678. Use the cleaning-indicator catalogue to maintain consistent product and holder identification across the monitoring programme.