MEDIWISH · Cleaning monitoring guide
Understand the parts of the cleaning test
“Test soil”, “cleaning strip” and “steel coupon” describe different parts of a cleaning test. Test soil is the challenge material to be removed. The card, strip or coupon is its carrier. A holder positions the carrier and can affect access of washing fluid. Some products supply these parts separately; others arrive as one complete device.
Prepared reagent and ready-to-use indicator
A powdered test-soil reagent is prepared and applied according to its own instructions and the assessment protocol. Preparation, application, conditioning and the chosen test surface need to be controlled for that method. A ready-to-use indicator already carries its applied soil, so it has a different preparation workflow.
Do not copy mixing ratios, drying times or reading criteria from another reagent or strip. An unspecified recipe cannot become a reproducible test merely by being called “artificial soil”. If the method requires a residual measurement after cleaning, identify that measurement separately from the visual appearance of the test object.
What the formats change
| Format | What to identify | MEDIWISH example |
|---|---|---|
| Powdered reagent | Preparation instructions, application surface and assessment method. | Powder Test Soil REF 7892398. |
| Polymer strip | Applied working area, specified holder and intended process. | Universal Strip REF 647913, holder 12-5814. |
| Steel strip or coupon | Carrier dimensions, test-soil area and holder. | CEI CVC-50, 30 × 6 × 0.65 mm, holder 12-5814. |
| Complete device | Integral holder and disposal of the whole assembly. | CVWC-20, supplied as 20 complete single-use devices. |
CEI CVC-50 is the existing MEDIWISH CEI model also identified as MZS-50.100. Rev05 specifies 50 individually packaged steel strips. The steel carrier does not turn it into a direct test for residual protein on an instrument. CVWC-20’s plastic holder is part of its single-use assembly, not an accessory to retain for replacement strips.
Soil colour and composition are separate questions
A red mark does not establish that a soil contains real blood. A black pattern does not establish a higher cleaning challenge. Terms such as synthetic, blood-containing or blood-free need support in documents for that exact REF. Do not extend a composition statement from one MEDIWISH model to the whole range.
The carrier and holder also matter when comparing results. Two products with a similar colour or shape cannot be assumed equivalent. Keep any comparison tied to the test conditions, endpoint and actual configurations assessed.
Choose the method that matches the protocol
Start with whether the procedure calls for applying a reagent or using a prepared indicator. Then verify the applicable process, holder, reading method and traceability requirements. ISO 15883-5:2021 provides a standards context for washer-disinfector cleaning-efficacy test methods; its mention does not certify any product.
Review reagent and steel-strip formats
Review Powder Test Soil REF 7892398 for reagent-based work, or CEI CVC-50 stainless steel cleaning indicators for the prepared steel-strip format. Request the current instructions relevant to your intended assessment.