Mediwish Helix Test Hollow Load Process Challenge Device

MZS-012-4 Daily Steam Penetration ControlTest (Helix Test) on a Benchtop Vacuum SteriliserCE approval Helix Process Challenge Device ( PCD ) Test

MEDIWISH® Helix Test System. Hollow Load PCD.REF

Directions for Use.  INDICATIONS FOR USE

Mediwish Helix Test Hollow Load Process Challenge Device (PCD) is compliant with EN 867-5 for pre-vacuum steam sterilization of hollow loads. Mediwish Helix Test hollow load PCD verifies air removal, deep vacuum achievement necessary for lumens, steam penetration, and exposure levels. It is a reusable device that can be utilized in each sterilization load, as an independent control device for decisions regarding load release or for daily release of pre-vacuum steam sterilizers should hollow load tests be required.

CRITICAL PARAMETERS (in a standard hospital steam sterilizer)

3.5 minutes at 134°C; 15 minutes at 121°C STEAM process.

PACKAGING

The Mediwish Helix Test can be reused for 200 cycles and consists of one Mediwish Helix Test PCD device 200 Mediwish Helix Test strip monitors. Indicator Strip is self-adhesive for easy documentation.

DESCRIPTION

The disassembled Mediwish Helix Test Process Challenge Device consists of the following parts: a standardized lumen connected to an end-cap, an indicator capsule with sealing “O' ring, Class 2 ink chemistry Conforms to ISO 11140-1, Class 2.

INSTRUCTIONS FOR USE

  1. Unscrew the indicator capsule from the lumen end cap.
    1. Remove a Mediwish Helix Test indicator from the protective pouch and fold it in half at the score line with the indicator ink facing inward.
    2. Insert it into the indicator capsule with the fold first.
      1. Replace the indicator capsule by twisting by hand until snug. NOTE: DO NOT over tighten the cap using tools.
      2. Place the assembled Mediwish Helix Test device in the sterilizer, in a horizontal position, near the coolest location in the sterilizer (usually on a shelf, near the drain, next to the door).
      3. Conduct the intended sterilizer cycle.
        1. At the end of the cycle, remove the Mediwish Helix Test PCD from the sterilizer and wait one minute to allow the device to cool before opening. CAUTION: The Mediwish Helix Test device may still be hot. Use hand protection if necessary.
        2. Open the indicator capsule and remove the Mediwish Helix Test monitor strip. Read the results.
        3. Document the results as required by hospital procedures.

10. Allow the device to cool and dry prior to next use. Look for one or more slugs of water that could block the free transmission of air or steam throughout the entire length of the lumen. Do not use if water slugs are present. Placing the open device in a warm dry place can accelerate drying.

11. The indicator strip has a self-adhesive backing and can be applied to the sterilization records.

INTERPRETATION OF RESULTS

Color change from Pink to Black indicates satisfactory air removal and steam penetration within the load. The presence of any pink color remaining indicates insufficient sterilization conditions and re-­sterilization is recommended. Follow hospital procedures for re-sterilization of the load.

REMOVING THE RELEASE LINER

Remove release liner by bending the folded monitor back on itself, which will cause the release liner to pull away from the monitor slightly at the fold line. Grasp the release liner and pull to remove. Adhere individual monitors to records as specified by hospital policy.

STORAGE

Store at room temperature 10°- 40°C. Do not remove Mediwish Helix Test monitors from protective foil pouch until ready for processing. Protect from moisture and excess humidity by resealing the foil bag after each use. Once processed, the monitor strips are stable and require no special storage conditions.

EXPIRY DATE

The expiry date is printed on the product packaging.

LOT NUMBER

A unique identification code, |LOT], is printed on each indicator and packaging labels.

INTERFERING SUBSTANCES OR CONDITIONS

There are NO KNOWN INTERFERING SUBSTANCES OR CONDITIONS that could affect the intended use of the indicator or adversely affect the indicator performance.

RELEASE OF TOXIC SUBSTANCES

The indicator releases NO KNOWN TOXIC SUBSTANCES in sufficient quantities to cause either a health hazard, or a hazard to the intended properties of the product being sterilized before, during or after the sterilization process.

Ordering Information

 

Art No.

Qty.

MZS -12

1 device + 200 indicators

Price
12 USD
| Contact Us Today
Your name: *
Phone number: *
Your message:
text
 
NEWS & EVENTS
0704.18

mediwish_participate_in_meditech_2018_clombia.jpgDear customers! Mediwish Co., Ltd will attend in MEDITECH 2018, BOGOTA, COLOMBIA. 3-6 July 2018. Pavillion 18-23, Booth No:2511. WELCOME TO US!

0704.18

mediwish_healthgb_exibition_england.jpgMEDIWISH participates in Health GB Exibition and Congress 2018, in MANCHESTER, ENGLAND. Welcome to visit our booth.

0704.18

mediwish_participate_cmef_2018.jpgDear customers! Mediwish Co., Ltd will attend one of the biggest and the most influential exibitions for medical market professionals during 11th-14th April 2018. Welcome to visit our stand. 

1601.17

Dear Customers! We wish you and yours family happy New Year!
Reminder, what our factory will close 23th Jan-04 Feb on holidays. WE will replay you after 04thFeb.

Adress:
#336 Tianhe Rd , Luyang Industrial Zone, Hefei, Anhui, China. Post code: 230041
Business hours:
from 9:00 to 18:00

mediwish_factory_address_map.jpg